Portable device for measuring blood test hemolyzed samples based on computer vision and neural network (2019)
- Authors:
- USP affiliated authors: ZANCUL, EDUARDO DE SENZI - EP ; DOMINGUES, PIETRO TERUYA - EP ; KAWAKAMI, LINCOLN MAKOTO - EP ; LOPES, KARYN MARTINELLI - EP
- Unidade: EP
- DOI: 10.1115/1.4043078
- Subjects: REDES NEURAIS; PROCESSAMENTO DIGITAL DE IMAGENS; PLASMA; SANGUE
- Agências de fomento:
- Language: Inglês
- Abstract: Hemolysis is a challenging problem and still represents a frequent source of errors in blood test laboratory practice. Due to the broad and heterogeneous bias induced in the measurement of several parameters by hemolysis, inaccurate results may be reported, and the patient may be required to repeat sample collection, delaying diagnosis. Existing automated laboratory devices including hemolysis detection are not suitable for lower volume and smaller sample collection sites. In many situations, hemolysis is still detected by visual inspection of the sample after centrifugation, during the blood test pre-analytical stage. Visual inspection is highly dependent on a qualified workforce, subjective to interpretation discrepancies, and thus difficult to standardize. The paper aims to describe the design and performance of a portable device for measuring hemolyzed samples based on computer vision and neural network. The results indicate that the device provides hemolysis indexes with sufficient accuracy to guide laboratory decision in the blood test pre-analytical stage.
- Imprenta:
- Publisher place: New York, NY
- Date published: 2019
- Source:
- Título: Journal of Medical Devices
- ISSN: 1932-6181
- Volume/Número/Paginação/Ano: v. 13, n.2, p. 021004-1 - 021004-8, Jun 2019
- Status:
- Artigo possui versão em acesso aberto em repositório (Green Open Access)
- Versão do Documento:
- Versão submetida (Pré-print)
- Acessar versão aberta:
-
ABNT
LOPES, Karyn Martinelli et al. Portable device for measuring blood test hemolyzed samples based on computer vision and neural network. Journal of Medical Devices, v. 13, n. 2, p. 021004-1 - 021004-8, 2019Tradução . . Disponível em: https://doi.org/10.1115/1.4043078. Acesso em: 02 abr. 2026. -
APA
Lopes, K. M., Silva, F. H. da, Maldonado, A. S. G., Santiago, S. A., Pires, T. A., Ferrer, C. M., et al. (2019). Portable device for measuring blood test hemolyzed samples based on computer vision and neural network. Journal of Medical Devices, 13( 2), 021004-1 - 021004-8. doi:10.1115/1.4043078 -
NLM
Lopes KM, Silva FH da, Maldonado ASG, Santiago SA, Pires TA, Ferrer CM, Mena SJ, Moura MEG, Domingues PT, Kawakami LM, Zancul E de S. Portable device for measuring blood test hemolyzed samples based on computer vision and neural network [Internet]. Journal of Medical Devices. 2019 ; 13( 2): 021004-1 - 021004-8.[citado 2026 abr. 02 ] Available from: https://doi.org/10.1115/1.4043078 -
Vancouver
Lopes KM, Silva FH da, Maldonado ASG, Santiago SA, Pires TA, Ferrer CM, Mena SJ, Moura MEG, Domingues PT, Kawakami LM, Zancul E de S. Portable device for measuring blood test hemolyzed samples based on computer vision and neural network [Internet]. Journal of Medical Devices. 2019 ; 13( 2): 021004-1 - 021004-8.[citado 2026 abr. 02 ] Available from: https://doi.org/10.1115/1.4043078 - Application of set-based concurrent engineering principles in R&D project prioritization
- Diferentes perspectivas de sucesso em projetos de inovação: uma análise comparativa das abordagens design for six sigma e design thinking
- Distribuição de propostas de projetos de inovação: fatores de decisão de alocação entre equipes de DFSS e Design Thinking
- Análise dos fatores influenciadores no desenvolvimento de projetos de soluções médicas em empresas empreendedoras.
- Análise da adequação da metodologia de desenvolvimento de produtos e serviços em projetos de inovação
- A história de um poka-yoke digital
- Concepção, implantação e aplicação de uma Fábrica de Aprendizagem voltada para a Indústria 4.0
- A preliminary analysis of product-service system (PSS) types and implementation
- Product-service system types and implementation approach
- It is a great honour for us to serve as the Scientific Advisory Board for the 1st Research Conference of the Design Factory Global Network (DFGN): introduction from Scientific Advisory Board. [Introdução]
Informações sobre a disponibilidade de versões do artigo em acesso aberto coletadas automaticamente via oaDOI API (Unpaywall).
Por se tratar de integração com serviço externo, podem existir diferentes versões do trabalho (como preprints ou postprints), que podem diferir da versão publicada.
Download do texto completo
| Tipo | Nome | Link | |
|---|---|---|---|
| Zancul-2019-portable_devi... |
How to cite
A citação é gerada automaticamente e pode não estar totalmente de acordo com as normas