Precipitation titrations using an automatic titrator based on a multicommutated unsegmented flow system (2000)
- Authors:
- USP affiliated authors: REIS, BOAVENTURA FREIRE DOS - CENA ; ZAGATTO, ELIAS AYRES GUIDETTI - CENA
- Unidade: CENA
- DOI: 10.1039/a907917c
- Subjects: QUÍMICA ANALÍTICA; QUÍMICA ANALÍTICA INSTRUMENTAL
- Language: Inglês
- Abstract: An automatic flow titrator based on a multicommutated unsegmented flow system was used to perform precipitation titrations with potentiometric detection. The titration approach is based on sequential introduction of increasing titrant volumes into a mixing chamber. The system allows the attainment of complete curves, the titration end point being determined by the Gran method, as is usually employed in batch potentiometric titrations. The theoretical model for the analytical process is presented and discussed and the determination of titrand concentration is performed without any prior calibration. The automatic flow titrator developed was applied to chloride determination in bottled natural waters and the results obtained were reproducible (RSD 3.2 and 3.8% for repeatability and reproducibility, respectively) and in good agreement with those given by the reference procedure (relative deviation within the range from -4.6 to 4.8%)
- Imprenta:
- Source:
- Título: The Analyst
- Volume/Número/Paginação/Ano: v. 125, p. 333-340, 2000
- Este periódico é de acesso aberto
- Este artigo NÃO é de acesso aberto
-
ABNT
ALMEIDA, C M N V et al. Precipitation titrations using an automatic titrator based on a multicommutated unsegmented flow system. The Analyst, v. 125, p. 333-340, 2000Tradução . . Disponível em: https://doi.org/10.1039/a907917c. Acesso em: 22 jan. 2026. -
APA
Almeida, C. M. N. V., Araujo, M. C. U., Lapa, R. A. S., Lima, J. L. F. C., Reis, B. F. dos, & Zagatto, E. A. G. (2000). Precipitation titrations using an automatic titrator based on a multicommutated unsegmented flow system. The Analyst, 125, 333-340. doi:10.1039/a907917c -
NLM
Almeida CMNV, Araujo MCU, Lapa RAS, Lima JLFC, Reis BF dos, Zagatto EAG. Precipitation titrations using an automatic titrator based on a multicommutated unsegmented flow system [Internet]. The Analyst. 2000 ; 125 333-340.[citado 2026 jan. 22 ] Available from: https://doi.org/10.1039/a907917c -
Vancouver
Almeida CMNV, Araujo MCU, Lapa RAS, Lima JLFC, Reis BF dos, Zagatto EAG. Precipitation titrations using an automatic titrator based on a multicommutated unsegmented flow system [Internet]. The Analyst. 2000 ; 125 333-340.[citado 2026 jan. 22 ] Available from: https://doi.org/10.1039/a907917c - Detecting and circumventing sources of inaccuracy in flow analysis
- Multicommutation in flow analysis . Part 4. Computer-assisted splitting for spectrophotometric determination of copper and zinc in plants
- Trial measurements in flow analysis
- Development of a potentiometric procedure for determination of glycerol plus 2,3 butanediol in wine by sequential injection analysis
- Desenvolvimento de um procedimento potenciométrico para a determinação de glicerol em aguardentes empregando sistema de análises em fluxo
- Exploitation of tandem streams for carry-over compensation in flow analysis: I. Turbidimetric determination of potassium in fertilizers
- Multi-pumping in flow analysis: concepts, instrumentation, potentialities
- Multi-task flow system for potentiometric analysis: its application to the determination of vitamin B-6 in pharmaceuticals
- Automated spectrophotometric determination of clomipramine on a multicommuted flow system
- Tandem flow analysis applied to food and agricultural samples
Informações sobre o DOI: 10.1039/a907917c (Fonte: oaDOI API)
How to cite
A citação é gerada automaticamente e pode não estar totalmente de acordo com as normas